Simulation Study to Improve Ion Transmission Efficiency in Mass Spectrometry with a Dual Ion Funnel Ion Source

Mass Spectrometry Letters / Mass Spectrometry Letters, (P)2233-4203; (E)2093-8950
2013, v.4 no.4, pp.91-94
https://doi.org/10.5478/MSL.2013.4.4.91
Baek Sun Jong (Korea Basic Science Institute)
Kim Seung Yong (Korea Basic Science Institute)
Kim Taeman (Bioneer Corporation)
Kim Hyun Sik (Korea Basic Science Institute)
  • Downloaded
  • Viewed

Abstract

We performed computer simulations to improve transmission efficiencies of a dual ion funnel system implemented onan FT-ICR MS. We found that the low m/z range from 50 to 150 could be significantly improved by operating the two ion funnelsat different RF amplitudes. These new operational conditions could be applied to analyze metabolome samples, whichrequire high sensitivity in the m/z range from 50 to 1,000.

keywords
FT-ICR, ion funnel, trajectory calculation, simulations, SIMION